A MedWatch Safety Alert was added to the FDA Recalls webpage.
TOPIC: Giraffe Incubators and OmniBeds by GE Healthcare: Class I Recall - Due to Potential for Infants to Fall
BACKGROUND: Incubators and warmers are used for infant care in a hospital setting. Incubators are small beds enclosed in clear, hard plastic that is temperature-controlled for infants who are unable to maintain their body temperature on their own. Infant warmers are small beds with heaters over them to warm infants who are unable to maintain their body temperature on their own. Healthcare providers can give care to the baby through holes in the sides of the incubator, also called portholes.
AUDIENCE: Patient, Pediatrics, Health Professional, Risk Manager
ISSUE: GE Healthcare is recalling the Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestations because the bedside panels can be upright and look closed but not be securely latched. The portholes can also look closed when not securely latched. If a canopy cover is used, it can hold the bedside panel or porthole door closed without being securely latched. If an infant comes in contact with a bedside panel or porthole that is unlatched, the panel or porthole can disengage and fall open, no longer protecting the infant from falling.
The firm has received 6 reports of infant falls with injuries such as skull fractures, hematoma and edema. No deaths were reported.
RECOMMENDATION: On November 8, 2019 GE Healthcare sent an updated letter to customers stating that users can continue to use their Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation systems by following these instructions:
- Every time the bedside panel is closed, users must pull on the bedside panel to make sure the bedside panel is latched, and the red tab is no longer visible (see Figures 1 and 2).
- Users must pull on the porthole doors every time the bedside panel or porthole is closed to make sure the porthole door is latched (see Figure 3).
- Apply the safety warning labels that were included with the letter to the Incubator or OmniBed using the instructions provided.
- Review the user manual Addendum and place it with the Incubator or OmniBed User Manual.
- Place all three (3) of the provided “Giraffe Incubator/OmniBed Risk of Patient Fall” posters in prominent locations for healthcare staff and ensure the posters are displayed for the lifetime of the Incubator or OmniBed(s).
- Confirm that the information from the customer letter and user manual Addendum are properly disseminated to all users that handle the Incubator or OmniBed.
- Ensure that all hospital staff who open panels or portholes or otherwise come into contact with the device understand these instructions.
- Ensure that any spare bedside panels or end panels that are not installed on incubator or OmniBed units are appropriately labeled according to the instructions. Additional labels can be obtained from a local GE representative.
- Complete and return the provided form to document that all hospital staff who come into contact with the incubator or OmniBed are trained on the proper closing and latching of the devices and that the steps provided in the instructions are followed.